CEL-63x Environmental Noise Meter
75 Casella CEL
without the wind screen, then you must set the calibration level to
113.88 dB. You must round this value to 113.9 dB.
o Calibrator level 113.98 dB
o Correction –0.1 dB
o Calibration level 113.88 dB rounded to 113.9 dB
If you use a Cirrus CR:515 calibrator and the certified output from the
calibrator is 94.04 dB, and you intend to use the instrument with the
wind screen fitted, then you must set the calibration level to 94.04 dB.
You must round this value to 94.0 dB.
o Calibrator level 94.04 dB
o Correction 0.0 dB
o Calibration level 94.04 dB, rounded to 94.0 dB
Refer to section 3.4 “Calibrating the instrument” on page 12 for instructions to
set the calibration level and calibrate the instrument.
9.2 Response characteristics
Table 7. Casella CEL-63X with CEL-251 Microphone Free Field response with and
without Wind Screen
0 Degree
Free Field
Response
0 Degree
Free Field
Corrections
0 Degree
Free Field
Response
with wind
Screen
0 Degree
Free Field
Corrections
with Wind
Screen
Expanded
uncertainty
of
Corrections
(k=2)